2. Acquisition channels: Maximum 9 channels can be supported for simultaneous acquisition, which is suitable for simultaneous recording of multiple neural nuclei or simultaneous recording of multiple animals. The system is equipped with low auto fluorescence fiber, which does not need to be bleached in advance to effectively reduce background fluorescence interference and achieve close multi-location recording.
3. System configuration: Integrated signal acquisition and digital signal synchronization module; configuration of 4 Input interfaces to support four types of external signal input automatic tagging; 4 Output interfaces to support the output TTL signal trigger external third-party devices to meet the closed-loop research. The host and computer are equipped with USB3.0 and 2.0 different interfaces; configured with integrated professional acquisition and analysis software, the software integrates behavioral recording and analysis module, which can realize automatic marking of animal tracking; the data analysis function can support synchronous comparison with other physiological signals. System shipment includes host, computer, software, behavior camera and support, low auto fluorescence fiber, black ceramic sleeve.
4. Compatible with Optogenetics: Equipped with a port for connecting optogenetics, support 635nm light input of the optogenetics to complete stimulation and neural signal recording at the same site
5. Acquisition camera: Highly sensitive CMOS detector; Dual detector +TDM designs; acquisition frequency up to 250fps; exposure time: 1-100ms; gain multiplier: 1-100.
6. Power adjustment: Excitation light power can be adjusted directly on the operating software, Min 0μW, Max≥100μW, adjustment range 0~100% display, adjustment accuracy 0.1μW.
7. Marking mode: The software has three marking functions, external marking, manual marking, behavior ROI area marking; the software supports the simultaneous setting of no less than 20 or more manual marking and automatic marking and customize the marking shortcut keys, names and colors.
8.Marking position editing and data interception function: You can delete the event marker or modify the marker color according to the actual experimental situation; you can choose the time period of interest to synchronously intercept fluorescence data and behavioral video, which can be played simultaneously.
9. Fluorescence mode: Software preset 6 fluorescence excitation output modes, which can be applied to different experimental application scenarios.
10. Acquisition mode: Acquisition mode can be set through the software, can customize the setting of acquisition time and other parameters, also can choose two modes of continuous acquisition and interval acquisition.
11. Fiber ROI settings: ROI can be adjusted by the external adjustment knob of the machine to adjust the fiber end-face imaging state. ROI settings of all channels can be adjusted according to the actual experiment by dragging or inputting values to synchronize the size.
12. Behavior acquisition settings: Behavior video and fluorescence data can be set in the same software interface for simultaneous acquisition. The behavioral video acquisition can support multiple behavioral cameras to record simultaneously, and the video frame rate (30-150FPS optional) and resolution (1920x1080 multiple optional) can be set directly.
13. Behavioral analysis area settings: Behavioral acquisition supports setting specific Area areas for behavioral analysis and simultaneously setting 9 ROI areas and naming them, 3 kinds of ROI drawing tools: circle, square, polygon, can be set when the animal enters a ROI area for automatic marking, and later can analyze behavioral data on time, location and other parameters.
14. Interaction mode: The acquisition stage can output TTL signal to trigger external devices when the fluorescence signal is greater than the set Δ f/f threshold or when the animal enters a certain ROI area, and parameters such as pulse width frequency (0-500Hz) of the output TTL signal, the delay and duration time can be set, and the TTL signal output can be tested directly.
15. Acquisition window: Different channel data are directly displayed on the window for easy direct identification and comparison; different channels of fluorescence data and different wavelengths of fluorescence data axes can be adjusted individually.
Statement of Need and Purpose:
This requirement aims to procure Tricolor multichannel fiber photometry system to allow researchers to detect the activity of neurons and behavior of neuronal populations in the neural circuit study. This requirement aims to procure tricolor multichannel fiber photometry system to enable researchers to take advantage of optogenetics, to explore detailed insight into the activity and behavior of neuronal populations. This technique stimulates neurons with light and measures fluorescence signals that correspond to calcium dynamics by detecting changes in the fluorescence intensity from a genetically encoded calcium indicator (GECI), calcium dynamics related to activities and patterns in neural circuits can be measured in real time. The technique can help in developing treatments for neurological diseases and brain-related traumatic injuries. Its use of light offers many advantages over other methods of neuromodulation, including a temporal resolution on the order of milliseconds, closely matching natural neuronal activity; and the stimulation of multiple neurons sets independently using different wavelengths.
Background and Objectives:
The National Institutes of Health (NIH) is the nation’s leading medical research agency and the primary Federal agency whose mission is to seek fundamental knowledge about the nature and behavior of living systems and the application of that knowledge. The mission of the NIH also includes enhancing health, lengthening life, reducing illness and disability, conducting, supporting, and making medical discoveries that improves and saves people’s health and lives.
The National Institute of Neurological Disorders and Stroke (NINDS) SNC group routinely uses different manipulations to modulate neuronal activities and behaviors of subpopulations to observe neural terminal construction between the presynaptic and postsynaptic connections, these observations will allow the researchers to study the neural circuits, neuronal synaptic transmission or excitability of target neurons in brain regions.
Quantity: 1 package including one permanent license, Data analysis software, recording software, computer laptop, recording system including camera.
Delivery Date: The Contractor shall deliver and install the equipment to the NIH Delivery Address in Bethesda, Maryland no later than 6-8 weeks after receipt of the order (ARO). The equipment shall be delivered and installed between the hours of 8:00am and 5:00pm EST, Monday through Friday. Delivery must be Freight on Board (FOB) Destination and include inside delivery. The equipment shall be delivered and installed in coordination with the NINDS personal [to be named at time of award] at the time the equipment is ready to be shipped.
Other requirements: The Contractor shall provide an Original Equipment Manufacturer (OEM) certified service technician to perform setup and installation.
Warranty - The Contractor shall warrant that the equipment will be free from defects for a period of at least 12 months from the date of installation, inspection by the Government, and acceptance.
Interested parties may identify in writing their interest and capability in response to this requirement. Responses to this notice shall contain sufficient information to establish the interested parties’ bona-fide capabilities for fulfilling the requirement and include: unit price, list price, shipping and handling costs, the delivery period after contract award, the prompt payment discount terms, the F.O.B. Point (Destination or Origin), the Dun & Bradstreet Number (DUNS), the Taxpayer Identification Number (TIN), and the certification of business size. All offerors must have an active registration in the System for Award Management (SAM) www.sam.gov.
All responses must be received by closing date and must reference the announcement. Responses may be submitted electronically to the attention of the contract specialist. Fax responses will not be accepted.
All responsible sources may submit a bid, proposal, or quotation which shall be considered by the agency.